The DAD Display Format command is only available when a DAD detector is selected on the Method Configuration screen.
The DAD Display Format command is only available when a DAD detector is selected on the Method Configuration screen.
Use this command to display and modify the DAD display format.
Use this command to display and modify the DAD display format.


'''Tool Bar Shortcut'''
'''Tool Bar Shortcut'''
Zeile 11: Zeile 12:
!'''Dialog Entries''' !!  
!'''Dialog Entries''' !!  
|-
|-
|style="width:100px;text-align:left; vertical-align:top;"|Display Format
|style="width:130px;text-align:left; vertical-align:top;"|'''Display Format'''
|style="width:525px;text-align:left; vertical-align:top;"|'''Absorbance Scale:''' Specifies a value of absorbance scale from a fixed set to draw the Contour and 3-D displays and the associated chromatogram displays on the Data Display window. The default value is Auto. The range of selections with the L-4500 or L-7450 detectors is either from 2.0 to 0.005 or from 0.2 to 0.0005 depending on the Absorbance Mode selected in DAD Setup (Normal or Low). The range of selection with the L-7455 detector is from 2.0 to 0.0005.
|style="width:525px;text-align:left; vertical-align:top;"|'''Absorbance Scale:''' Specifies a value of absorbance scale from a fixed set to draw the Contour and 3-D displays and the associated chromatogram displays on the Data Display window. The default value is Auto. The range of selections with the L-4500 or L-7450 detectors is either from 2.0 to 0.005 or from 0.2 to 0.0005 depending on the Absorbance Mode selected in DAD Setup (Normal or Low). The range of selection with the L-7455 detector is from 2.0 to 0.0005.


'''Offset:''' The value specified is used to set the zero position on the AU scale for drawing Contour and 3-D displays. The default value is 0.0%.Time Range: Specifies the time range used to draw the Contour and 3-D displays.Wavelength Range: Specifies the wavelength range used to draw the Contour and 3-D displays.
'''Offset:''' The value specified is used to set the zero position on the AU scale for drawing Contour and 3-D displays. The default value is 0.0%.Time Range: Specifies the time range used to draw the Contour and 3-D displays.Wavelength Range: Specifies the wavelength range used to draw the Contour and 3-D displays.
|-
|-
|style="width:100px;text-align:left; vertical-align:top;"|Spectrum Display
|style="width:130px;text-align:left; vertical-align:top;"|'''Spectrum Display'''
|style="width:525px;text-align:left; vertical-align:top;"|'''Absorbance:''' Select to plot the spectrum using Absorbance Scale specified in Display Format.
|style="width:525px;text-align:left; vertical-align:top;"|'''Absorbance:''' Select to plot the spectrum using Absorbance Scale specified in Display Format.


Zeile 25: Zeile 26:
'''Auto BG Subtraction:''' If you are including spectra in a report, select to include background subtracted spectra, automatically.  
'''Auto BG Subtraction:''' If you are including spectra in a report, select to include background subtracted spectra, automatically.  
|-
|-
|style="width:100px;text-align:left; vertical-align:top;"|3-D Mesh Display Options
|style="width:130px;text-align:left; vertical-align:top;"|'''3-D Mesh Display Options'''
|style="width:525px;text-align:left; vertical-align:top;"|
|style="width:525px;text-align:left; vertical-align:top;"|
'''Resolution:''' Select one from Fine, Medium, or Coarse.
'''Resolution:''' Select one from Fine, Medium, or Coarse.
Zeile 33: Zeile 34:
'''Orientation:''' Set Title and Rotation angles. Defaults are 50 and 30, respectively. Check Mirror View to display the mirror image.
'''Orientation:''' Set Title and Rotation angles. Defaults are 50 and 30, respectively. Check Mirror View to display the mirror image.
|-
|-
|style="width:100px;text-align:left; vertical-align:top;"|Report Spectra
|style="width:100px;text-align:left; vertical-align:top;"|'''Report Spectra'''
|style="width:525px;text-align:left; vertical-align:top;"|Report Spectra is disabled when Chromatograms to Create is set to None on DAD Data Processing screen. Add to check mark to select Peak Top and Sides (3 spectra) or Peak Top Only (1 spectra).
|style="width:525px;text-align:left; vertical-align:top;"|Report Spectra is disabled when Chromatograms to Create is set to None on DAD Data Processing screen. Add to check mark to select Peak Top and Sides (3 spectra) or Peak Top Only (1 spectra).
|-
|-
|}
|}
[[File:L29.png]]
[[File:L29.png]]

Aktuelle Version vom 9. August 2018, 12:02 Uhr

The DAD Display Format command is only available when a DAD detector is selected on the Method Configuration screen. Use this command to display and modify the DAD display format.


Tool Bar Shortcut


Click on DAD Display Format.png.

Dialog Entries
Display Format Absorbance Scale: Specifies a value of absorbance scale from a fixed set to draw the Contour and 3-D displays and the associated chromatogram displays on the Data Display window. The default value is Auto. The range of selections with the L-4500 or L-7450 detectors is either from 2.0 to 0.005 or from 0.2 to 0.0005 depending on the Absorbance Mode selected in DAD Setup (Normal or Low). The range of selection with the L-7455 detector is from 2.0 to 0.0005.

Offset: The value specified is used to set the zero position on the AU scale for drawing Contour and 3-D displays. The default value is 0.0%.Time Range: Specifies the time range used to draw the Contour and 3-D displays.Wavelength Range: Specifies the wavelength range used to draw the Contour and 3-D displays.

Spectrum Display Absorbance: Select to plot the spectrum using Absorbance Scale specified in Display Format.

Normalized: Select to plot the spectrum using the normalized scale where the maximum intensity of each spectrum is scaled to 1.0.

Auto Mark Peak WL: Select to display labels on peaks detected on the spectrum plot.

Auto BG Subtraction: If you are including spectra in a report, select to include background subtracted spectra, automatically.

3-D Mesh Display Options

Resolution: Select one from Fine, Medium, or Coarse.

Display: Check Spectra Only to display the 3-D as a series of spectra.

Orientation: Set Title and Rotation angles. Defaults are 50 and 30, respectively. Check Mirror View to display the mirror image.

Report Spectra Report Spectra is disabled when Chromatograms to Create is set to None on DAD Data Processing screen. Add to check mark to select Peak Top and Sides (3 spectra) or Peak Top Only (1 spectra).