La Chrom HPLC D-7000: Unterschied zwischen den Versionen

Aus Hergipedia
Zur Navigation springen Zur Suche springen
Keine Bearbeitungszusammenfassung
Keine Bearbeitungszusammenfassung
 
(12 dazwischenliegende Versionen desselben Benutzers werden nicht angezeigt)
Zeile 217: Zeile 217:


== Measuring ==
== Measuring ==
After purging the system and creating a method the column you want to use should be chosen and flushed for about ~30 min with the desired solvent mixture. Columns don´t like harsh polarity changes, therefor a slow change of the solvent polarity to the desired solvent mixture is preferred. After conditioning the column, the prepared sample has to be put into the sample table of the autosampler. Then the position and the related method are listed in the sequence table.
After purging the system and creating a method, the column you want to use should be chosen and flushed for about ~30 min with the desired solvent mixture. Columns don´t like harsh polarity changes, therefore a slow change of the solvent polarity to the desired solvent mixture is preferred. After conditioning the column, the prepared sample has to be put into the sample table of the autosampler. Then the position and the related method are listed in the sequence table.


===Creating a sequence table===
===Creating a sequence table===
Zeile 278: Zeile 278:
|- style="vertical-align:top;"
|- style="vertical-align:top;"
|style="width:400px;text-align:; vertical-align:top;"|The easiest way to flush the column is manually.  
|style="width:400px;text-align:; vertical-align:top;"|The easiest way to flush the column is manually.  
*Manually set the pump (Manual set [[File:Manual set.png|30px]]; 0[[File:Zahl 2.png |30px]]; 5 [[File:Zahl 5.png |30px]]; Enter [[File:Enter.png|30px]]; 4 [[File:Zahl 2.png |30px]]; 5 [[File:Zahl 5.png |30px]]; Enter [[File:Enter.png|30px]]; 0[[File:Zahl 2.png |30px]]; 5 [[File:Zahl 5.png |30px]]; Enter [[File:Enter.png|30px]]; Enter [[File:Enter.png|30px]]; Enter [[File:Enter.png|30px]]; Enter [[File:Enter.png|30px]])
*Manually set the pump (Manual set [[File:Manual set.png|30px]]; 0 [[File:Zahl 0.png |30px]]; 5 [[File:Zahl 5.png |30px]]; Enter [[File:Enter.png|30px]]; 4 [[File:Zahl4.png |30px]]; 5 [[File:Zahl 5.png |30px]]; Enter [[File:Enter.png|30px]]; 0 [[File:Zahl 0.png |30px]]; 5 [[File:Zahl 5.png |30px]]; Enter [[File:Enter.png|30px]]; Enter [[File:Enter.png|30px]]; Enter [[File:Enter.png|30px]]; Enter [[File:Enter.png|30px]])
*Turn on the Pump (Pump ON/OFF [[File:ON OFF.png|30px]]);  
*Turn on the Pump (Pump ON/OFF [[File:ON OFF.png|30px]]);  
[[File:spüleinstellung.png|top|center|thumb|500px|''Flushing'']]
[[File:spüleinstellung.png|top|center|thumb|400px|''Flushing'']]
Let it run for about 30 min.
Let it run for about 30 min.
*Turn the pump off(Pump ON/OFF [[File:ON OFF.png|30px]]).
*Turn the pump off(Pump ON/OFF [[File:ON OFF.png|30px]]).
Zeile 289: Zeile 289:
*Switch of the pump (Pump ON/OFF [[File:ON OFF.png|30px]]).
*Switch of the pump (Pump ON/OFF [[File:ON OFF.png|30px]]).


|style="width:525px;text-align:left; vertical-align:top;"| [[File:.png|top|center|thumb|500px|''"Sequence table"'']] [[File:60_Start.png|top|center|thumb|500px|''"Sequence table"'']]
|style="width:525px;text-align:left; vertical-align:top;"| [[File:Pumpe L 7100.png|top|center|thumb|500px|Pumpe]] [[File:Tastatur.png|top|center|thumb|500px|key panel]]


|- style="vertical-align:top;"
|style="width:400px;text-align:; vertical-align:top;"|Im Folgenden ruft man das ''"Sequence Parameter"'' Menu auf '''(5)'''. Dort kann man einstellen das sobald er mit der Messung der Sequenz fertig die UV-Lampe und die Pumpe ausgeschaltet wird '''(6)'''. Diese Einstellung bestätigt man mit OK '''(7)''' und startet diese '''(8)'''.
|style="width:525px;text-align:left; vertical-align:top;"| [[File:61_Start.png|top|center|thumb|500px|''"Sequence Parameter"'']][[File:62_Start.png|top|center|thumb|500px|Messung]][[File:63_Start.png|top|center|thumb|500px|Lauf]]
|}
|}


===Schließen===
===Shut Down===
{| class="wikitable"
{| class="wikitable"


! Erklärung !! Bild
! Explanation !! Picture
|- style="vertical-align:top;"
|- style="vertical-align:top;"


|style="width:400px;text-align:; vertical-align:top;"|Ist die Sequenz durchgelaufen, öffnet sich das Reportfenster und die Pumpe und die UV-Lampe schalten sich aus. Nun kann das Reportfenster geschlossen werden '''(1)'''
|style="width:400px;text-align:; vertical-align:top;"|Turn the pump off.


|style="width:525px;text-align:left; vertical-align:top;"|[[File:64_Start.png|top|center|thumb|500px|Lauf]]
|style="width:525px;text-align:left; vertical-align:top;"|[[File:Shut5.png|top|center|thumb|500px|Pumpmenu]][[File:Shut4.png|top|center|thumb|500px|Press OK]]




|- style="vertical-align:top;"
|- style="vertical-align:top;"
|style="width:400px;text-align:; vertical-align:top;"| Da die Pumpe und die UV-Lampe ausgeschalte sind (grau) '''(2,3)''', kann nun das Programm geschlossen werden '''(4)'''.  
|style="width:400px;text-align:; vertical-align:top;"| Dissconect HPLC.
|style="width:525px;text-align:left; vertical-align:top;"|[[File:66_Start.png|top|center|thumb|500px|Programm schließen]]
|style="width:525px;text-align:left; vertical-align:top;"|[[File:Shut1.png|top|center|thumb|500px|]] [[File:Shut2.png|top|center|thumb|500px|]] [[File:Shut3.png|top|center|thumb|500px|]]


|- style="vertical-align:top;"
|style="width:400px;text-align:; vertical-align:top;"| Switch off the HPLC.
|style="width:525px;text-align:left; vertical-align:top;"|[[File:L HPLC Nummeriert.png|top|center|thumb|500px|]]


|- style="vertical-align:top;"
|style="width:400px;text-align:; vertical-align:top;"| Die Nachfrage ob man das Programm wirklich schließen will kann einfach bestätigt werden '''(1)''', ebenso die Frage ob die Pumpe und die Lampe aus ist '''(2)''', falls diese grau unterlegt sind.
|style="width:525px;text-align:left; vertical-align:top;"|[[File:67_Start.png|top|center|thumb|500px|Programm schließen]] [[File:68_Start.png|top|center|thumb|500px|Programm schließen]][[File:69_Start.png|top|center|thumb|500px|Programm schließen]]
|}
|}



Aktuelle Version vom 7. September 2018, 11:31 Uhr

!!!Use the operator book!!! DO NOT FORGET!!!!

Everybody who is using the HPLC should enter his name, the uesed colum, the solvent mixture and the back pressure of the system in the user book.

Assembly

A HPLC consists of different parts. The La Chrom HPLC D-7000 in room 118 is made up of the folowing modules

La Chrom HPLC analytisch
Module Product Number Picture Manual
Solvent bottles
solvent bottles
Pump L7100
Quaternary Pump
Quaternary Pump
Autosampler/ Injector L-7200
Autosampler
Autosampler
Interface D-7000
Interface
Interface
DAD-Detector L-7455
DAD-Detector
DAD-Detector
RI-Detector L-7490
RI-Detector
RI-Detector
Column Switcher BESTA-High-Pressure-Motorvalve
Column Switcher
Column Switcher
Column Oven L-7350
Column Oven
Column Oven
Solvent Waste
Solvent Waste
Computer (Windows 2000)
Computer

Preparations

Column

The HPLC is connected to a column switcher, which gives you the opportunity to choose between five different columns. If the column you want to use isn´t included you have to change one of the columns. The method you want to use should be compatible with the conditions the column can stand (pH, pressure, solvent etc.). After the installation you should take care, that all connections are tight. All columns can be tempered in the column oven.

Solvent

* Before starting a measurement, the solvent level should be checked and if necessary refilled. Don´t forget to check the Autosampler solvent bottle. *All used solvents have to be degassed and be compatible with the used column. Especially the RI-Detector won´t work with not proper degassed solvents. * It is required, that all solvents have HPLC quality and are particle free. Small particles can permanently block the capillaries and valves. Avoid the use of the following steel-corrosive solvents: #Solutions of alkali halides and their respective acids (for example, lithium iodide, potassium chloride, and so on). #High concentrations of inorganic acids like sulfuric acid, especially at higher temperatures (replace, if your chromatography method allows, by phosphoric acid or phosphate buffer which are less corrosive against stainless steel). #Chromatographic grade ethers, which can contain peroxides (for example, THF, dioxane, diisopropylether) such ethers should be filtered through dry aluminium oxide which adsorbs the peroxides. #Halogenated solvents or mixtures, which form radicals and/or acids, for example: ::2CHCl3 + O2 → 2COCl2 + 2HCl ::This reaction, in which stainless steel probably acts as a catalyst, occurs quickly with dried chloroform if the drying process removes the stabilizing alcohol. Prevent Blocking of Solvent Filters Contaminated solvents or algae growth in the solvent bottle will reduce the lifetime of the solvent filter and will influence the performance of the pump. This is especially true for aqueous solvents or phosphate buffers (pH 4 to 7). The following suggestions will prolong lifetime of the solvent filter and will maintain the performance of the pump: * Use sterile, if possible amber, solvent bottles to slow down algae growth. * Filter solvents through filters or membranes that remove algae. * Exchange solvents every two days or refilter. * If the application permits add 0.0001–0.001 M sodium azide or 0.1 % TFA to the solvent. * Avoid exposure of the solvent bottles to direct sunlight. * When using buffer solutions, flush the system with water before switching it off. NOTE Never use the system without solvent filter installed.

Waste

Before the use of the HPLC, the waste container should be checked and emptied if necessary.

Sample

The sample preparation is an important part of the HPLC analysis. It is important, that the sample is dissolved and filtered before the measurement. This prevents the contamination of the column and the precipitation of the sample on the column bed. The sample has to be dissolved at any time and shouldn´t precipitated from the eluent. Therefor it is not thoughtful to dissolve the sample in pure methanol or acetonitrile, if you are working with an RP-phase column. The risk of sample precipitation during the measurement is to high. Sample precipitation causes a rising of the back pressure and clogging of the column, which leads to an irreversible poor resolution.

What is the right solvent mixture to dissolve many sample?

This question can´t be answered easily and depends on the sample. The most important thing is, that the sample is in solution during the whole measurement. How can you find the right solvent mixture? One way is to do this in practice is the following: At first take 1 mg of your sample and dissolve it in 500 µL HPLC Methanol respectively ACN. Is the solution still clear you can successively add water in 100 µL steps. After each addition check if the solution is still clear. If the solution is clear go on with addition until you reach a total volume of 1000 µL. If it becomes cloudy stop the addition of water and add 100 µL of HPLC Methanol respectively ACN. The precipitated sample should dissolve again. Then calculate the solvent ratio. This is the maximum solvent ratio you can use for the eluent without the risk of precipitation of the sample on the column bed. Solvent mixtures with higher amount of water would lead to precipitation. To get rid of remaining particles, the sample solution is filtered with a syringe filter.

Start

Bevor you start the HPLC it is important to make preparations (see previous paragraph).

Explanation Picture
If you want to start the HPLC you have to switch on the different parts of the HPLC-assembly. # Pump # Interface/ Interface # Autosampler/ Injector # RI-Detector # DAD-Detector # Column Oven # Column Switcher # Computer (if it is not already running)
Startreihenfolge
Open the programm "D 7000- HSM" by double clicking the shortcut with the left mouse button.
Desktop
Programm
Programm
After the programm has opened click the white " i " (1) on blue ground. A window opens and shows you the hardware status. For connecting the computer with the HPLC click on " Initialize" (2). The status changes to "Downloading" (3). After a while the HPLC and computer are connected and the status changes to "Idle" (3a). All connected parts are shown on the left (4). The connection then can be confirmed with "OK" (5).
Start
Initializing
Initializing
Connected
After the HPLC is connected to the computer you have to purge the system. To purge it manually, is the easiest way to do it. *Manually set the pump (Manual set Manual set.png; 2 Zahl 2.png; 5 Zahl 5.png; Enter Enter.png; 2 Zahl 2.png; 5 Zahl 5.png; Enter Enter.png; 2 Zahl 2.png; 5 Zahl 5.png; Enter Enter.png; Enter Enter.png; Enter Enter.png; Enter Enter.png) *Turn on the Pump (Pump ON/OFF ON OFF.png); *Open purge valve (1); *Start purging (Purge2 Purge.png); *Purge the system for about 10 minuntes; *Stop purging (Purge2 Purge.png); *Close purge valve (2); *Turn the pump off(Pump ON/OFF ON OFF.png).
Purge valve
When you are finished purging the system you have to choose the applications of a certain operator. In our case "Allgemeinheit Lindhorst" (1, 2, 3).
Initializing
Initializing

Method creating/editing

To be able to separate substances with the HPLC it is nesassary to setup a method. The easiest way to creat a method is by editing an existing method.

Explanation Picture
Click on "Method Setup" and choose a method you want to edit (1,2,3,4)
load Method
load Method
When the window opens, the Method Information Icon is selectedMethod Information.png and the window shows the general Method information(1). These infomation can be changed to the desired ones. The upper part of the window shows different Icons (2). Each Icon leads you to a different method parameters, which you can change to the wished ones. Module Setup Menu *Method Information.png Method Information Use this command if you want to review or modify general Method information such as the Method title, Method name, or solvent names. *Method Configuration.pngMethod Configuration Use this command if you want to review or modify Method Configuration parameters. *Pump Setup.pngPump Setup Use this command if you want to review or modify pump parameters or the solvent table. *Autosampler Setup.png Autosampler Setup Use this command if you want to review or modify autosampler parameters. *Channel 1-DAD Setup.pngChannel 2-RI Setup.png Channel 1 or 2 Detector Setup Use these commands if you want to review or modify the parameters of the detectors on channel 1 or channel 2. *Event Signal.png Event Signal Setup Use this command to initiate the review or modification of Event Signal parameters. *Time Summary.png Time Summary Use this command to display the pump (solvent) table and event table graphically. Data Process Setup Menu *Channel 1.pngChannel 2.png Channel 1 or Channel 2 Use these commands to select a channel (1 or 2) for the subsequent review or modification of data processing parameters. *Calculation Method.png Calculation Method Use this command to review or modify the calculation method. *Component Table.png Component Table Use this command to review or modify the Component Table. *Integration Time Table.png Integration Time Table Use this command to review or modify the Integration Time Table. *Data Processing.png DAD Data Processing Use this command to review or modify DAD data processing parameters. *Chromatogram Display Format.png Chromatogram Display Format Use this command to review or modify parameters. *DAD Display Format.png DAD Display FormatUse this command to review or modify DAD display format parameters. *Confiedence Report.png Confidence Report Use this command to review or modify confidence report parameters. *Report Format.png Report Format Use this command to review or modify report format parameters. *Update Format.png Update Method Use this command to update all parameters of current Method to all other open windows.
Methode laden
If you don´t want to go to much into Details editing the method. Here are the most relevant parameters marked, that should be adapted. *Method Information.png Method Information
L16 S.png
*Method Configuration.png Method Configuration
L17 S.png
*Pump Setup.png Pump Setup
L18 S.png
*Channel 1-DAD Setup.pngChannel 2-RI Setup.png Channel 1 or 2 Detector Setup
L20 S.png
*DAD Display Format.png DAD Display Format
L29 S.png
After you have changed all parameters to your own desires, you have to save the method. Click File (1) and Save Method As (2). In the opening Window you have to give the method a name (3), assign a Aplikation to it (4) and press OK (5).
Saving Method
Saving Method

Measuring

After purging the system and creating a method, the column you want to use should be chosen and flushed for about ~30 min with the desired solvent mixture. Columns don´t like harsh polarity changes, therefore a slow change of the solvent polarity to the desired solvent mixture is preferred. After conditioning the column, the prepared sample has to be put into the sample table of the autosampler. Then the position and the related method are listed in the sequence table.

Creating a sequence table

Explanation Picture
to start creating a new Sample Table click (1, 2, 3).
Probentisch
Sequence table
Change the sample table parameters as desired and switch to the Sample table (1).
Sample table parameter
Fill in the Sample table.
Sample Table
Save Sample Table.(1, 2, 3)
Saving Sample Table

Start Measurment

Explanation Picture
Turn on the pump. (1,2).
S8.png
S9.png
Open the sample table to aquire data (1,2).
Start
open sample table
Start the measurment (1).Be patient the system is not so fast !!!!!!!!!!
Start Measurment

Flushing

After the last measurement the column has to be flushed and stored at a certain solvent mixture. The column is normally flushed with a solvent mixture, which has a higher organic content then the solvent used during the measurement.

Explanation Picture
The easiest way to flush the column is manually. *Manually set the pump (Manual set Manual set.png; 0 Zahl 0.png; 5 Zahl 5.png; Enter Enter.png; 4 Zahl4.png; 5 Zahl 5.png; Enter Enter.png; 0 Zahl 0.png; 5 Zahl 5.png; Enter Enter.png; Enter Enter.png; Enter Enter.png; Enter Enter.png) *Turn on the Pump (Pump ON/OFF ON OFF.png);
Flushing
Let it run for about 30 min. *Turn the pump off(Pump ON/OFF ON OFF.png). *Change the pump manually to the storange solvent mixture (this depend on the column); *Turn the pump back on (Pump ON/OFF ON OFF.png) Let the system run for additional 10 min *Switch of the pump (Pump ON/OFF ON OFF.png).
Pumpe
key panel

Shut Down

Explanation Picture
Turn the pump off.
Pumpmenu
Press OK
Dissconect HPLC.
Shut1.png
Shut2.png
Shut3.png
Switch off the HPLC.
L HPLC Nummeriert.png

Auswerten

Die Auswertung kann entweder an Hand des ausgedruckten Reports oder an Hand eines im offline Programm von Chemstation modifizierten Reports erfolgen. Von Interesse sind hierbei insbesondere die Retentionszeiten und Integrale der einzelnen Signale, auch das UV-Spektrum der einzelnen Signale gibt wichtige Hinweise.

Abschließende Hinweise

Vor Verwendung der HPLC ist es sinnvoll seine Zielsetzung klar zu definieren. Es bietet sich auch an für eine gute Dokumentation bei jeder Messungen ein Laufzettel zu verwenden, in dem alle Parameter der Messung vermerkt sind um eine Reproduzierbarkeit zu gewährleisten. Ein Vordruck kann hier oder im Desktopbereich des HPLC-Computers gefunden werden und bei Bedarf ausgedruckt werden.